<!DOCTYPE html> UHV Magnetron Sputtering System — L.V. Kirensky Institute of Physics
You are here: Home / Institute / Scientific equipment / UHV Magnetron Sputtering System

UHV Magnetron Sputtering System

The UHV magnetron sputtering system comprises sputtering and loading chambers. The sputtering chamber contains four magnetron sources.

Сверхвысоковакуумная система магнетронного напыления с тремя мишенями

Specification

The sputtering chamber is evacuated with a rate of 500 l/s by turbomolecular, volute, and titanium sublimation pumps. The loading chamber is evacuated with a rate of 60 l/s by turbomolecular and volute pumps. The system contains three 500W dc generators and a 300W high-frequency generator.

A substrate heater automatically rotated with a speed of 0–20 rpm ensures heating up to 600°С. An SHQ-15 PID controller provides temperature stability within ±1°С. A 100W generator supplies rf voltage to a substrate.

The system is equipped with

  • a MICROVISION quadrupole mass analyzer (1–200 amu) with a resolution of < 5% and a working pressure of 10-4 –10-13 mbar;
  • a film thickness meter:
  • film thickness measurement accuracy 1A;
  • sampling frequency of 4 GHz;
  • time resolution 0.1/s,
  • stores data on nine materials.
Filed under:
« March 2025 »
March
MoTuWeThFrSaSu
12
3456789
10111213141516
17181920212223
24252627282930
31